You can attach the Ozone Calibration Source(OCS™) output directly to the inlet of any ozone monitor; the excess flow is vented through an ozone scrubber internal to the OCS. Although the instrument comes calibrated against a NIST-traceable ozone standard, the menu allows the user to change the calibration parameters of the ozone output if desired. The scheduler function allows you to schedule an autocalibration once daily that executes a series of up to 10 programmable output steps.
The Ozone Cal Source™ is housed in the same small, rugged instrument case as the Model 202 and Model 205 Ozone Monitors™.
Some applications of the Model 306 Ozone Calibration Source™ include:
- Calibration of ozone monitors in the field. This highly portable transfer standard is especially useful for maintaining the calibrations of networks of ozone monitors.
- Testing of materials such as rubber, plastics and paint for effects of ozone exposure.
- Studies of the effect of ozone on plants where a calibrated source of ozone is required.
Method of Ozone Production | UV Photolysis of O2 at 185 nm |
Ozone Concentration Range | 0 ppbv and 30 ppbv to 1,000 ppbv |
Precision and Accuracy of Output |
Greater of 2 ppbv or 2% of ozone concentration |
Response Time for Change in Ozone Output Concentration |
30 s to reach 95% of concentration change |
Flow Rate | 3.0 Liter/min volumetric |
Power Requirements | 12 V dc or 120/240 V ac, 18 watt |
Size | 3.5 x 8.5 x 11 inches (9 x 21 x 29 cm) |
Weight | 5.6 lb (2.6 kg) |
Analog Output | Yes; voltage output tracks ozone output and provides system error codes |
Serial Connection | Yes; for remote operation and system diagnostics |
Programmable Output | Yes; user-specified ozone output for up to 10 steps from 0 to 1,000 ppbv; stored in internal memory |
Scheduler | Yes; once-daily execution of the step program at user-specified hour |
Baud Rate | 4800 |